ACCU-CAL150 eng

UV Tester

ACCU-CAL150

ACCU-CAL150
HM-2 is a cumulative intensity, peak intensity, and light meter capable of measuring real-time illumination.
Feature
– UV UV energy through gyeonghwagi (J / cm ²) and UV
– UVA, UVA2, UVB, UVC, UVV can select only one of
– cumulative intensity, the peak intensity can be measured
SPECIFICATION
GENERAL SPECIFICATION
ITEM DESCRIPTION
 Suggested Operation Ranges *Standard High Range : UVA, UVB, UVV-100mW/cm2 to 10W/cm2/ UVC – 10mW/cm2 to 1W/cm2
*Mid-Range : UVA, UVB, UVV – 10mW/cm2 to 1W/cm2 / UVC : 1mW/cm2 to 100mW/cm2
*Low Power : UVA, UVB, UVV – 1mW/cm2 to 100mW/cm2 / UVC – 1mW/cm2 to 100mW/cm2
*Units will “turn on” and display data at irradiance values much lower than the suggested Operating Ranges. The suggested Operating Ranges are where the instrument performs best.
 Accuracy +/- 10%; +/- 5% typical
Spectral Ranges
(UV Power Puckⓡ II)
4-channel continuous monitoring. Standard version: 320-390nm(UVA), 280-320nm (UVB), 250-260nm (UVC), 395-445nm (UVV) / UVA2 Version:380-410nm (UVA2 replaces the UVC band)
Spectral Ranges
(UVICUREⓡ Plus II)
1-channel continuous monitoring. 320-390nm (UVA), 380-410nm(UVA2 for LED monitoring and additive bulb monitoring),280-320nm (UVB), 250-260nm(UVC), 395-445nm(UVV)
Spatial Response Approximately cosine
 Operating Temperature  0-75°C Internal temperature; tolerates high external temperatures for short periods(audible alarm indicates when temperature has exceeded tolerance)
 Time-Out Period 2 minutes DICPLAY mode (no key activity). A no time-out mode can be activated by EIT-IM.
Battery  Two user-replaceable AAA Alkaline Cells
 Battery Life Approx, 20 hours with display on
Dimensions 4.60 * 0.50 inches; 117mm * 12.7mm (D*H)
 Weight 10.1 ounces (289 grams)
 Instrument Materials  Aluminum, stainless steel
 Carrying Case Material  Cut polyurethane interior, scuff resistant nyoln exterior cover
 Carrying Case Weight 9 ounces (260 grams)
 Carrying Case Dimensions 10.75 * 3.5 * 7.75 inches; 274 * 89 * 197mm (W * H * D)
Feature & Advantage
– UV UV energy through gyeonghwagi (J / cm ²) and UV
– UVA, UVA2, UVB, UVC, UVV can select only one of
– cumulative intensity, the peak intensity can be measured
Detail
p46
– UVA (320~390nm)
– UVC (250~260nm)
– UVA2 (375~415nm)
– UVV (395~445nm)
– UVB (280~320nm)